Julabo FP89-HL Ultra-Low Refrigerated-Heating Circulator
$30,240.00 – $31,040.00
Julabo FP89-HL Ultra-Low Refrigerated-Heating Circulator
Superior models for most demanding applications
JULABO Ultra-Low Refrigerated Circulators for heating and cooling are suitable for external temperature tasks. Powerful circulating pump systems and high heating and cooling capacities guarantee short heat-up and cool-down times. ACC ‘Active Cooling Control‘ provides active cooling control across the whole temperature range. An energy saving feature with only minor loss of heat is provided by the proportional cooling capacity control in the ‘FP’ refrigeration units. The inevitable dust accumulation that occurs in refrigeration systems is solved by means of an easily removed venting grill, which allows for easy cleaning. The units are equipped with handles or castors for an easy transportation. All models have a drain tap on the front allowing the fluid to be drained easily. Additionally, the instruments are equipped with an improved insulation to help avoid ice-formation, and have a visual liquid level display.
Your advantages
-
- VFD COMFORT DISPLAY
-
- LCD DIALOG DISPLAY backlit for convenient interactive operation
-
- Keypad for setpoints, warning/safety values and menu functions
-
- ICC (Intelligent Cascade Control), self-optimizing temperature control
-
- TCF Temperature Control Features to optimize the control behaviour
-
- ATC3 3-Point-Calibration
-
- Pt100 External sensor connection for measurement and control
-
- SMART PUMP, electronically adjustable pump stages
-
- Early warning system for low liquid level (DBGM 203 06 059.8)
-
- Adjustable high temperature cut-out, visible via display
-
- RS232/RS485 interface for online communication
-
- Integrated programmer for 6 x 60 program steps
-
- Connections for solenoid valve and HSP booster pump
-
- Proportional cooling control
-
- Active Cooling Control
-
- Proportional cooling control
Order No. | 9312689 |
---|---|
Model series | High Tech |
Category | Ultra-Low Refrigerated-Heating Circulators |
Working temperature range (°C) | -90 … +100 |
Temperature stability (°C) | ±0.02 |
Setting / display resolution | 0.01 °C |
Integrated programmer | 6×60 steps |
Temperature Display | VFD , LCD |
Heating capacity (kW) | 1.3 |
Cooling capacity (Medium Ethanol) | °C 20/0/-20/-40/-60/-80 —- kW 1/0.92/0.88/0.75/0.58/0.2 |
Pump capacity flow rate (l/min) | 22-26 |
Pump capacity flow pressure (psi) | 5.8-10.2 |
Pump capacity flow suction (psi) | 2.9-5.8 |
Bath opening / bath depth (W x L / D inch) | 5.1 x 5.9 / 6.3 |
Pump connections | M16x1 |
Barbed fittings diameter (inner dia. / mm) | 8 / 12 |
Filling volume (liters) | 8 |
Refrigerant stage 1 | R404A |
Filling volume refrigerant stage 1 (g) | 500 |
Global Warming Potential for | 3922 |
Carbon dioxide equivalent stage 1 (t) | 1.961 |
Refrigerant stage 2 | R508B |
Filling volume refrigerant stage 2 (g) | 270 |
Global Warming Potential for | 13396 |
Carbon dioxide equivalent stage 2 (t) | 3.617 |
External Pt100 sensor connection | integrated |
Digital interfaces | RS232 RS485 Optional Profibus |
Ambient temperature | 5…40 °C |
Dimensions W x L x H (inch) | 21.7 x 23.6 x 36.2 |
Weight (LBS) | 297 |
Classification according to DIN12876-1 | Classification III (FL) |
Included with each unit | 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). |
Cooling of compressor | Air |
Power requirement V / Hz / A | 230/60/16 Additional voltage versions available |
Characteristics
-
- A perfect view
-
- Additional plain text information
-
- Pump stage and liquid level
-
- Comfortable and detailed
-
- For perfect results
‘Intelligent Cascade Control’, automatic & self optimizing adjustment of PID control parameters, temperature stability ±0.005 °C internal,
-
- Full control
-
- Highest measuring accuracy
-
- Energy saving cooling
-
- Consistent cooling capacity
-
- 100 % Cooling capacity
-
- Condensation and ice protection
-
- Clever pump system
-
- Control from the external application
-
- Connection compliant to standard
-
- Optimal program control
-
- Automatic control of operating time
-
- Early warning system for low liquid level
-
- Early warning system for high/low temperature limits
-
- Enhanced protective functions
-
- For flammable bath fluids
Optional: Electronic module with analog connections for external programming, temperature recorder, standby, alarm (Order no. 8 900 100).
Technical changes without prior notification.
Images may deviate from the original.
Specification: Julabo FP89-HL Ultra-Low Refrigerated-Heating Circulator
|