1 SciSolLogoWhite

(844) 404-4588

24/7 Customer Support

1200C Controlled Atmosphere Muffle Furnace ETL/CE Certified

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Kanthal elements, Mitsubishi refractory lining, Shimaden 40-seg PID controller, overheat protection, multiply inlet/outlets, PC interface, casters. ETL certified, standard.

$7,990.00$20,539.00

SKU: 694 Category:

 

 

 

GCF1200 series controlled atmosphere furnaces feature a chamber with Mitsubishi (Japan) high quality alumina fiber insulation and the famous Kanthal (Sweden) heating elements. A double walled steel internal housing helps minimize heat loss to exterior surface. Furnace operation is controlled by Shimaden (Japan) 40-segment digital controller with built-in RS485 digital communications port and USB adaptor, allowing the user to connect to a PC for remote control and monitoring of the furnace. You can also save or export test results.

The benefits of a controlled atmosphere furnace

An atmosphere furnace provides the perfect environment for various procedures in many laboratories, providing the controlled temperature and atmosphere specific projects require. The controlled atmosphere is desired to complete tasks that require exact temperatures with no room for error. This is the best way to ensure the successful outcome of your project because it is sealed tightly to ensure the proper environment.

Gases

One of the main components of the atmosphere furnace is the correct mixture of gases. The precise measurement of each gas is essential to the success of your project. You can use a separate chamber to mix the gases, which are then pumped into the furnace chamber where the work will be completed when they are done mixing. It is essential that the separate chamber that holds the gases does not receive any outside air because it could throw off the gas mixture and render your project useless.

Temperature

Proper temperature control is vital to the success of your project as well. Temperatures that are too high or too low will cause problems with the process and could cost your company a significant amount of money. Across International furnaces come with built-in 28-segment PID digital temperature controller for accurate monitoring and consistency.

Features

  • Multiple gas inlets and outlets with stainless steel ball valves. Flow meter to control inbound and outbound gas flows. Outlets can be ignited to burn off gas.
  • Furnace chamber is welded and oil leak tested to ensure air tightness. Furnace door is sealed with high temperature silicon gasket.
  • Built-in stainless steel water-cooling panel protects door sealing gasket when furnace is under heavy use.
  • Microprocessor based self-tuning PID control provides optimum thermal process with minimal overshoot.
  • Multiple atmosphere processes in a single cycle are possible (ex: binder burn out in air and parts sintering under rough vacuum or inert gas environment.)
  • Built-in ammeter and dual voltmeters for easy monitoring and troubleshooting.
  • Built-in computer interface.
  • Long life type K thermocouple.
  • ETL and CE certified, standard

Safety

  • Overheat protection shuts down the furnace if temperature is outside of acceptable range (refer to controller’s manual) or when the thermocouple is broken or malfunctions.
  • Power failure protection resumes furnace operation right after the point of failure when power is reestablished.

 

* Ultimate vacuum level for GCF furnaces is around 100 Pa with all flange connections, and 1000 Pa with hose barb connections. Please keep in mind that our GCF atmosphere furnaces are NOT designed to be vacuum furnaces, the vacuuming part is mainly for users to pre-vac the chamber, before putting in inert or other protective gases.
 

Specifications

 Electrical requirements  208-240V, 50/60Hz, single phase
 Heating element type
 Kanthal (Sweden) A1 2.0 resistance coil wire
 Min/max/constant temperature  Ambient/1200°C/1100°C
 Temperature controller

 Shimaden fp93 (Japan) with 4 patterns and
 40 segments (ie. 4 x 10 segments or 2 x 20 segments)
.

 Maximum heating rate  15°C / min
 Furnace cavity  Mitsubishi (Japan) high quality 1500 grade polycrystalline alumina fiber
 Thermocouple type   K
 Gas inlet/outlet  1/4″ hose barb with 1/4″ BSPT (British Standard Pipe Thread)
 Positive pressure  Not to exceed 0.02 MPa / 150 torr / 3 psi
 Temperature controller precision    +/- 1°C
 Working environments  Air, N2, Ar, NO2, CO, CO2, He, Ne, rough vacuum, etc. (Hydrogen, up to 5%)
 Certification
 ETL tested to UL and CSA standards, CE
 Download  Product brochure

 
Size comparison

 Chamber size (DxWxH)
 12x8x8″  16x12x12″  20x16x16″
 Chamber capacity (cu ft)  0.44  1.33  2.96
 Usable chamber space (DxWxH)  10x7x7″  14x11x11″  18x15x15″
 Maximum output  4 kW 19 amp
 9 kW 41 amp
 15 kW 69 amp
 Shipping dimensions (LxWxH)  38x26x44″  40x30x48″  49x44x61″
 Net weight (Lb)  350  400  800
 Shipping weight (Lb)  420  470  950

 

 
How to Program Across International Furnaces

 

 

 

 
Standard Package

Part description    

   

  Quantity  


Part image 

 GCF1200 controlled atmosphere furnace
 1 pc

 

 Type K thermocouple
 1 pc

 Flow meter
 1 pc


 Alumina cushion block  1 pc

 Furnace door block  1 pc

 Stainless steel tongs
 1 pc

 Thermal gloves  1 pair

 Spare fuse 2 pcs

 USB/RS485 computer interface kit  1 set

 Water pump  1 pc

 Furnace operation manual  1 pc


 Shimaden fp93 temperature controller manual & software   1 pc


Product Enquiry